66PACIFIC HOME

CALCULATOR PAGES
CO2 PAGES
PH PAGES
Simplest pH Meter
pH Test Kits
pH Probe
pH Calibration Buffers
pH Buffering with CaCO3

The Simplest Possible pH Meter




Building the Simplest Possible pH Meter

You can build a simple pH meter with just an inexpensive operational amplifier ("op amp") IC (a TL082 Dual JFET-input IC, about $2.00), 2 batteries, a digital volt meter, and a pH probe.

Schematic Diagram
Here's a schematic diagram of the circuit. Schematic diagram for the simplest possible pH meter. (Click to view full size.)Two 9-volt batteries power a high-input-impedance operational amplifier, such as a TL082. The pH probe is connected to the non-inverting input. The output voltage (V out), which is directly proportional to pH, is read with a voltmeter.

How It Works
The TL082 (or any other operational amplifier with a high impedance input) works as a unity gain buffer between the high impedance of the pH probe and the digital multimeter.

View of the Circuit
Here's a view of the entire circuit. Simplest possible pH meter - The TL082 IC is soldered directly to the BNC panel connector.The TL082 is mounted directly to a BNC panel connector.

Pictorial View of the Circuit
Here's a view of the entire circuit. Pictorial view of the simplest pH meter circuit. (Click to view full size.)This diagram shows the physical layout of the circuit. The voltmeter is connected to ground and pins 6 and 7 of the TL082 op amp, which are shorted together. In this circuit, pins 1, 2, and 3 of the IC are not used.



TL082 input pin soldered directly to the BNC panel connector Close-Up View
This circuit is so small and so simple that no circuit board is required. The prototype is soldered directly to the BNC panel connector.

Calibration and Use
In theory, a pH probe produces about 59 millivolts (mV) per pH unit, and at pH 7 (neutral pH) the probe produces 0 volts. Acid pHs produce negative voltages. Basic pHs produce positive pHs. For example, the system shown here reads +7.6 mV with the inputs shorted (due to the input offset error of the op amp). With the probe in pH 7.01 calibration buffer, the voltage is +4.6 mV. With pH 4.00 calibration buffer, the output voltage is 168.8 mV. This gives a range of 168.8 - 4.6 = 164.2 mV for 3.01 pH units, or 54.55 mV per pH unit.

If, for example, I measure 100 mV, the pH is 100 mV / 54.55 mV = 1.83 pH units deviation from 7.01, so I subtract 1.83 from 7.01 for a pH of 5.18. Creating a graph or a programming a calculator to do the arithmetic makes the process of determining pH simple, if not quite as simple as reading it off a pH meter's display.

Graphing the results
Here's a graph of pH and voltage from the calibration of the setup described here. Graph of pH and voltage (Click to view full size.)